Product Catalog

Mindray ePM10 Transport Patient Monitor

MDA Device Registration Number:
GC10656422-104738

Product Description

Compact, rugged transport monitor compliant with EN1789 and IEC 60601-1-12 for in-hospital and ambulance use. Features a 10.1" capacitive touchscreen, Platinum™ MPM parameter integration, 0.75m drop protection, and up to 6 hours battery runtime. Wireless connectivity allows real-time data viewing at the nurse station or any bedside monitor on the network.

Specifications

  • Weight: 3.2 kg (standard config, excl. recorder, battery, accessories)

  • Dimensions: 271 × 226 × 173 mm

  • Display: 10.1" capacitive touchscreen, 1280×800 px, up to 8 waveform channels

  • Drop protection: 0.75m (all 6 surfaces), EN1789 compliant

  • Patient types: Adult, Paediatric, Neonatal

ECG

  • Lead sets: 3 / 5 / 6 / 12-lead (auto recognition)

  • Bandwidth: Diagnostic 0.05–150 Hz | Monitor 0.5–40 Hz | Surgical 1–20 Hz

  • CMRR: >90 dB (diagnostic) | >105 dB (monitor/surgical/ST)

  • HR range: Adult 15–300 bpm | Paediatric/Neonate 15–350 bpm | Accuracy ±1 bpm or ±1%

  • Arrhythmia: 25 classifications, multi-lead

  • ST range: ±2.5 mV | Accuracy ±0.02 mV or ±10%

  • QT/QTc analysis: Bazett, Fridericia, Framingham, Hodges formulas

  • Defib protection: 5,000V / 360J, recovery <5s

  • Pace detection: ±2 mV–±700 mV amplitude; 0.1–2 ms width

SpO₂ (Mindray / Nellcor / Masimo options)

  • Range: 0–100% | Accuracy: ±2% adult/paediatric, ±3% neonate (70–100%)

  • PI (Perfusion Index): Yes (Mindray) | PR range: 20–300 bpm

NIBP

  • Technique: Oscillometry | Modes: Manual, Auto, STAT, Sequence

  • Systolic: Adult 25–290 / Paediatric 25–240 / Neonate 25–140 mmHg

  • Accuracy: Max mean error ±5 mmHg | Max SD 8 mmHg

Temperature

  • 2 channels | Range: 0–50°C | Accuracy: ±0.1°C | Resolution: 0.1°C

IBP

  • 2 channels | Range: −50–360 mmHg | Accuracy: ±1 mmHg or ±2% | PPV supported | ICP supported

C.O. (Cardiac Output)

  • Thermodilution | Range: 0.1–20 L/min | Accuracy: ±0.1 L/min or ±5%

CO₂ (optional — Artema sidestream, Oridion Microstream, or Capnostat mainstream)

  • Range: 0–150 mmHg | awRR: 0–150 rpm

Respiration

  • RR range: 0–200 rpm | Accuracy: ±1 rpm (0–120) | Apnea alarm: 10–40s

Data Storage (16G)

  • Trends: 240 hrs @ 1 min | 2,400 hrs @ 10 min

  • Events: 2,000 | NIBP: 3,000 sets | Full disclosure: 48 hrs | OxyCRG: 400 events

Power

  • AC: 100–240 VAC, 50/60 Hz

  • Battery: Li-ion 5,600 mAh smart battery

  • Runtime: ≥6 hrs (1× 5,600 mAh) | Charge: 5 hrs to 90%

Connectivity

  • Wi-Fi: IEEE 802.11a/b/g/n | WPA/WPA2 Enterprise | AES/TKIP encryption

  • Ports: RJ45, 2× USB 2.0, VGA, analog output/nurse call/defib sync, DC-in/docking

  • Barcode scanner: 1D and 2D supported

Environment

  • Operating: 0–40°C | Storage: −30–70°C | Humidity: 15–95% operating

Clinical Tools

  • ST Graphic™, EWS, GCS, 24h ECG summary, NIBP analysis, drug/hemodynamic/oxygenation/ventilation/renal calculations, titration table


Related Products

WhatsApp