Mindray ePM10 Transport Patient Monitor
MDA Device Registration Number:
GC10656422-104738
Product Description
Compact, rugged transport monitor compliant with EN1789 and IEC 60601-1-12 for in-hospital and ambulance use. Features a 10.1" capacitive touchscreen, Platinum™ MPM parameter integration, 0.75m drop protection, and up to 6 hours battery runtime. Wireless connectivity allows real-time data viewing at the nurse station or any bedside monitor on the network.
Specifications
Weight: 3.2 kg (standard config, excl. recorder, battery, accessories)
Dimensions: 271 × 226 × 173 mm
Display: 10.1" capacitive touchscreen, 1280×800 px, up to 8 waveform channels
Drop protection: 0.75m (all 6 surfaces), EN1789 compliant
Patient types: Adult, Paediatric, Neonatal
ECG
Lead sets: 3 / 5 / 6 / 12-lead (auto recognition)
Bandwidth: Diagnostic 0.05–150 Hz | Monitor 0.5–40 Hz | Surgical 1–20 Hz
CMRR: >90 dB (diagnostic) | >105 dB (monitor/surgical/ST)
HR range: Adult 15–300 bpm | Paediatric/Neonate 15–350 bpm | Accuracy ±1 bpm or ±1%
Arrhythmia: 25 classifications, multi-lead
ST range: ±2.5 mV | Accuracy ±0.02 mV or ±10%
QT/QTc analysis: Bazett, Fridericia, Framingham, Hodges formulas
Defib protection: 5,000V / 360J, recovery <5s
Pace detection: ±2 mV–±700 mV amplitude; 0.1–2 ms width
SpO₂ (Mindray / Nellcor / Masimo options)
Range: 0–100% | Accuracy: ±2% adult/paediatric, ±3% neonate (70–100%)
PI (Perfusion Index): Yes (Mindray) | PR range: 20–300 bpm
NIBP
Technique: Oscillometry | Modes: Manual, Auto, STAT, Sequence
Systolic: Adult 25–290 / Paediatric 25–240 / Neonate 25–140 mmHg
Accuracy: Max mean error ±5 mmHg | Max SD 8 mmHg
Temperature
2 channels | Range: 0–50°C | Accuracy: ±0.1°C | Resolution: 0.1°C
IBP
2 channels | Range: −50–360 mmHg | Accuracy: ±1 mmHg or ±2% | PPV supported | ICP supported
C.O. (Cardiac Output)
Thermodilution | Range: 0.1–20 L/min | Accuracy: ±0.1 L/min or ±5%
CO₂ (optional — Artema sidestream, Oridion Microstream, or Capnostat mainstream)
Range: 0–150 mmHg | awRR: 0–150 rpm
Respiration
RR range: 0–200 rpm | Accuracy: ±1 rpm (0–120) | Apnea alarm: 10–40s
Data Storage (16G)
Trends: 240 hrs @ 1 min | 2,400 hrs @ 10 min
Events: 2,000 | NIBP: 3,000 sets | Full disclosure: 48 hrs | OxyCRG: 400 events
Power
AC: 100–240 VAC, 50/60 Hz
Battery: Li-ion 5,600 mAh smart battery
Runtime: ≥6 hrs (1× 5,600 mAh) | Charge: 5 hrs to 90%
Connectivity
Wi-Fi: IEEE 802.11a/b/g/n | WPA/WPA2 Enterprise | AES/TKIP encryption
Ports: RJ45, 2× USB 2.0, VGA, analog output/nurse call/defib sync, DC-in/docking
Barcode scanner: 1D and 2D supported
Environment
Operating: 0–40°C | Storage: −30–70°C | Humidity: 15–95% operating
Clinical Tools
ST Graphic™, EWS, GCS, 24h ECG summary, NIBP analysis, drug/hemodynamic/oxygenation/ventilation/renal calculations, titration table